Leakage Current Tester According to IEC 60601-1, IEC60990, IEC60335, and other Safety Standards
DESCRIPTION
To ensure the safe use of electrical products, electrical safety tests such as insulation resistance, withstand voltage, ground resistance, and leakage current must always be conducted. Hioki Leakage Current Testers comply with laws and standards regarding medical electrical equipment as well as non-medical electrical equipment, and can be used to measure leakage current in all types of electrical products from computers to medical equipment.
The ST5540 consists of measurement networks that simulate the human body and a high-frequency volt meter, and is also capable of switching power supply polarities and making measurements under simulated single fault conditions in equipment under test.
Key Features
• Compliance with IEC 60601-1:2005 Ed 3.0, JIS T 0601-1:2012 for medical-use electrical devices and essential to electrical safety (*Starting on June 1, 2012, medical electrical equipment sold in the EU must comply). Model ST5540 comply with IEC 60601-1:2005+ A1:2012 (Ed 3.1), and IEC 62353 of 2017
• Compliance with Electrical Appliances and Materials Safety Act, JIS, IEC, and UL standards for general-use electrical devices
• Uninterrupted polarity switching function dramatically reduces cycle time
• Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards
• Touch panel features simple, interactive operation
• Communications functionality and external I/O support allow automatic testing on production lines
MODEL NO. (ORDER CODE)
ST5540 | for medical-use electrical devices |
---|
Note: The photograph portrays the Japanese version of the product. Other countries’ versions of the product lack the outlet on the front panel. Note: Always use an isolation transformer when measuring leak current for medical-use electrical devices. The ST5540 does not include an isolation transformer. When measuring medical-use electrical devices, use a step-up isolation transformer or similar component operating at 110% of the rated supply voltage as the power supply for the device under test.
LEAK CURRENT MEASUREMENT – ESSENTIAL TO ELECTRICAL SAFETY
Compliance with IEC 60601-1:2005 (3rd Edition) is now mandatory.
The ST5540 also complies with JIS T0601-1:2012.
The ST5540 series features an improved measuring method and dramatically faster cycle times thanks to its uninterrupted polarity switching capability. The new devices support rated currents of up to 20A, making it more than ideal for use with products built to new standards.
LEAK CURRENT TESTING OF MEDICAL ELECTRICAL DEVICES
In order to prevent the danger of electric shock, electrical devices use power supplies that are isolated from parts of the device that may come into contact with the body. However, it is impossible to achieve infinite insulation resistance. Some leak current always exists, and its magnitude changes as the insulation degrades over time. The LEAK CURRENT HiTESTER ST5540/ST5541 provides an easy-to-operate solution for measuring leak current in electrical devices, making it eminently suitable for use in an extensive array of applications, ranging from production lines to equipment maintenance and inspections.
COMBINE WITH MODEL 3157 TO MEET JIS T 1022:2006: SAFETY REQUIREMENTS OF ELECTRICAL INSTALLATIONS FOR MEDICALLY USED ROOMS IN HOSPITALS AND CLINICS
Combine the ST5540 with Hioki Model 3157 AC Grounding HiTester for Leak Current and Safety Conductor Testing
The following are key parts of any safety inspection of electrical equipment:
•Leak current test: Measure with the ST5540 and ST5541.
•Safety conductor test (also known as a ground line resistance test or ground conductor test): Measure with the 3157.
The 3157 can also be used for conducting measurements under the JIS T 1022:2006 safety standard for hospital electrical equipment.
BASIC SPECIFICATIONS (ACCURACY GUARANTEED FOR 1 YEAR, POST-ADJUSTMENT ACCURACY GUARANTEED FOR 1 YEAR)
Measurement methods | Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground. |
---|---|
Measurement modes | Leak current measurement, voltage measurement, safety conductor current measurement |
Standards compliance (NW: Body simulated resistance) | [NW-A] • Electrical Appliances and Materials Safety Act [NW-B1] • Medical electrical equipment: IEC 60601-1:1988+ A1:1993+ A2:1995, JIS T 0601-1:1999 [NW-B2] • Medical electrical equipment: IEC 60601-1:2005+ A1:2012, JIS T 0601-1:2012 and complement 1:2014, IEC 62353 [NW-C] • Measurement of touch current and protective conductor current: IEC 60990:2016 • Electrical equipment for measurement, control, and laboratory use: IEC 61010-1:2010+ A1:2016 • Information technology equipment: IEC60950-1:2005+ A1:2009+ A2:2013 • Audio, video and similar electronic apparatus: IEC 60065:2014 • Personnel Protection Systems for EV: UL 2231-1:2012 (Amended 2016), UL-2231-2:2012 (Amended 2016) [NW-D] • For UL: UL 1492:1996 (Amended 2013) [NW-G] • Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1:2010+ A1:2016 |
Leak current measurement | Ground leak current, 3 types of contact current, 7 types of patient leak current, patient measurement current, 4 types of total patient leak current, free current measurement, 3 types of enclosure leak current |
Measurement current | DC, AC (true rms, 0.1 Hz to 1 MHz), AC+DC (true rms, 0.1 Hz to 1 MHz), AC peak (15 Hz to 1 MHz) |
Measurement ranges | DC / AC / AC+DC mode: 50.00 mA/ 5.000 mA/ 500.0 μA/ 50.00 μA AC peak mode: 75.0 mA/ 10.00 mA/ 1.000 mA/ 500.0 μA |
Measurement accuracy (current measurement) | DC measurement: ±2.0% rdg. ±6 dgt. (typ.) AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.) AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.) |
Interfaces | External I/O, medical device relay output, USB 1.1 (communications), RS-232C |
Functionality | 110% voltage application, automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc. |
Power supply | 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power |
Target device power supply input | 100 to 250 V AC, 50/60 Hz Rated current input from terminal block: 20 A |
Target device power supply output | Output from terminal block: 20 A Output from outlet: 15 A |
Dimensions and mass | 320 mm (12.60 in)W × 110 mm (4.33 in)H × 253 mm (9.96 in)D, 4.5 kg (158.7 oz) |
Accessories | Test lead L2200 (for ST5540, Red ×2, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, User guide ×1, CD-ROM ×1 |
CATALOGS
CATALOG: LEAK CURRENT HITESTER ST5540, ST5541
USER GUIDES
LEAK CURRENT HITESTER ST5540
DECLARATION OF CONFORMITY
LEAK CURRENT HITESTER ST5540,ST5541 [100KB]
DRIVERS, FIRMWARE
ST5540 ST5541 MEMORY DATA RETRIEVING SOFTWARE
LABVIEW DRIVER FOR LEAK CURRENT HITESTER ST554X SERIES
USB DRIVER (DT4200SERIES 3800SERIES 7016 3554 ST5540 ST5541 BT4560)
SAMPLE PROGRAM FOR COMMUNICATION WITH HIOKI PRODUCTS
ST5540, ST5541 INSTRUCTION MANUAL
OPTIONS
Connection Options
* L9217: for motor analysis input
CONNECTION CABLE 9444
9 pin – 9 pin straight, 1.5 m (4.92 ft) length
Input cable (G)
* For the MR8990 *Voltage is limited to the specifications of the input modules in use.
TEST LEAD L2200
70 cm (2.30ft) length, detachable large alligator clips or needle tips are bundled, CAT IV 600V, CAT III 1000V
PC communication
*The 9151-02 is only for the RM3545-01
RS-232C CABLE 9637
For the PC, 9 pin – 9 pin, cross, 1.8 m (5.91 ft) length
Printer options
*PRINTER UNIT is a built-in option that must be specified upon order.
RECORDING PAPER 1196
For the Printer 9442, 112 mm (4.41 in) × 25 m (82.03 ft), 10 rolls/set
PRINTER 9442
For printing numerical values 112 mm (4.41 in) paper width, not CE marked
AC ADAPTER 9443-02
For the Printer 9442, EU type
Test probes
ENCLOSURE PROBE 9195
For the ST5540 series, 3156/3155