d. IMPEDANCE ANALYZER IM7585

d. IMPEDANCE ANALYZER IM7585

d. IMPEDANCE ANALYZER IM7585

1MHz to 1.3GHz Measurement Frequency Impedance Analyzer with 0.5ms Test Speed and 0.07% Variability

DESCRIPTION

Hioki LCR Meters and Impedance Analyzers range from 1mHz to 3GHz devices to suit a wide range of applications in the testing of electronic components. The IM7585 Impedance Analyzer offers a top measurement time of 0.5ms over a 1MHz to 1.3GHz frequency range and superior stability, making it ideal for R&D as well as high volume production of ferrite chip beads and chip inductors.

KEY FEATURES

• 1 MHz to 1.3 GHz testing source frequency
• Fastest test speed of 0.5 msec (Analog measurement time)
• 0.07% measured value variability (when measuring at 1GHz)
• ±0.65% rdg. basic accuracy
• Half-rack size body and palm-sized test head
• Comprehensive contact check (via DCR testing, Hi-Z reject or waveform judgment)
• Make frequency sweeps, level sweeps and time interval measurements in Analyzer Mode

MODEL NO. (ORDER CODE)

IM7585-01Connection cable 1 m is bundled
IM7585-02Connection cable 2 m is bundled

Note: The instrument does not ship with a test fixture or probe. A test fixture designed specifically for use with the Impedance Analyzer is required.

LAN
USB 2.0
CE
GPIB
RS232C
https://youtube.com/watch?v=B_pNUmtSxq8%3Frel%3D0%26showinfo%3D0%26enablejsapi%3D1%26wmode%3Dopaque
https://youtube.com/watch?v=Hi6-VG8skAg%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

IMPEDANCE ANALYZER IM7580 SERIES:LCR MEASUREMENT USING THE SMD TEST FIXTURE IM9201

The IM7580 series encompasses five models covering measurement frequencies of 1 MHz to 3 GHz. Used in combination with the Test Fixture IM9201, which can accommodate six SMD sizes, IM7580 series instruments allow you to measure specimens easily and reliably.

https://youtube.com/watch?v=nulxr0gUmLg%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

IMPEDANCE ANALYZER IM7580 SERIES:COMPARATOR FUNCTION

A comparator function lets you check whether measured values fall within a user-defined judgment range. This functionality is ideal for use in generating PASS/FAIL judgments for specimens.

https://youtube.com/watch?v=2bro2eovtzU%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

IMPEDANCE ANALYZER IM7580 SERIES:CONTACT CHECK FUNCTION

The instruments’ contact check function lets you check contact between the specimen and measurement terminals to detect faulty contact or verify good contact.

https://youtube.com/watch?v=XVcaB0MxOc4%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

IMPEDANCE ANALYZER IM7580 SERIES:HOW TO CONNECT THE TEST HEAD

Connect the cable of the test head to the impedance analyzer.
Tighten the nut until the handle of the torque wrench bends slightly.
Do not over tighten.

APPLICATION INFORMATION

Download the sample application and view the Communications Command User Manual.

Click here for the IMPEDANCE ANALYZER application.

ACCURACY CALCULATION INFORMATION

Calculate accuracy by simply entering a numerical value.

Click here to calculate accuracy.

BASIC SPECIFICATIONS (ACCURACY GUARANTEED FOR 1 YEAR, POST-ADJUSTMENT ACCURACY GUARANTEED FOR 1 YEAR)

Measurement modesLCR mode, Analyzer mode (sweeps with measurement frequency and measurement level), Continuous measurement mode
Measurement parametersZ, Y, θ, Rs (ESR), Rp, X, G, B, Cs, Cp, Ls, Lp, D (tanδ), Q
Measurable range100 mΩ to 5 kΩ
Display rangeZ: 0.00 m to 9.99999 GΩ / Rs, Rp, X: ± (0.00 m to 9.99999 GΩ)
Ls, Lp: ± (0.00000 n to 9.99999 GH) / Q: ± (0.00 to 9999.99)
θ: ± (0.000° to 180.000°), Cs, Cp: ± (0.00000 p to 9.99999 GF)
D: ± (0.00000 to 9.99999), Y: (0.000 n to 9.99999 GS)
G, B: ± (0.000 n to 9.99999 GS), Δ%: ± (0.000 % to 999.999 %)
Basic accuracyZ: ±0.65 % rdg. θ: ±0.38°
Measurement frequency1 MHz to 1.3 GHz (100 kHz setting resolution)
Measurement signal levelPower: -40.0 dBm to +1.0 dBm
Voltage: 4 mV to 502 mVrms
Current: 0.09 mA to 10.04 mArms
Output impedance50 Ω (at 10 MHz)
Display8.4-inch color TFT with touch screen
Measurement speedsFAST: 0.5 ms (Analog measurement time, typical value)
FunctionsContact check, Comparator, BIN measurement (classification), Panel loading/saving, Memory function, Equivalent circuit analysis, Correlation compensation
InterfacesEXT I/O (Handler), USB communication, USB memory, LAN, RS-232C (optional), GP-IB (optional)
Power supply100 to 240 V AC, 50/60 Hz, 70 VA max.
Dimensions and massMain unit: 215 mm (8.46 in) W × 200 mm (7.87 in) H × 348 mm (13.70 in) D, 8.0 kg (282.2 oz)
Test head: 90 mm (3.54 in) W × 64 mm (2.52 in) H × 24 mm (0.94 in) D, 300 g (10.58 oz)
AccessoriesTest head ×1, Connection cable ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1, Power cord ×1

CATALOGS

SMD TEST FIXTURE IM9201 

DOWNLOAD

COVER MEASUREMENT FREQUENCIES FROM 100 KHZ TO 3 GHZ, CHOOSE FROM 5 MODELS: IM7580 SERIES 

DOWNLOAD

IMPEDANCE ANALYZER IM7580 SERIES 

DOWNLOAD

USER GUIDES

LCR APPLICATION GUIDE 

DOWNLOAD

DECLARATION OF CONFORMITY

IMPEDANCE ANALYZER IM7583-01,IM7583-02,IM7585-01,IM7585-02 [122KB] 

DOWNLOAD

OPTIONS

PC communication

*The 9151-02 is only for the RM3545-01

GP-IB INTERFACE Z3000

RS-232C INTERFACE Z3001

GP-IB CONNECTOR CABLE 9151-02

2 m (6.56 ft) length

RS-232C CABLE 9637

For the PC, 9 pin – 9 pin, cross, 1.8 m (5.91 ft) length

Probe and Test fixtures

SMD TEST FIXTURE IM9201

• Direct connection two-terminal measurement type • DC to 3 GHz • Measurable sample sizes: 0201 to 1210 (EIA)

TEST FIXTURE STAND IM9200

Includes magnifying glass

CALIBRATION KIT IM9905

Open/Short/Load set

ADAPTER(3.5mm/7mm) IM9906

3.5 mm (0.14 in) male to 7 mm (0.28 in) conversion