DESCRIPTION

Hioki LCR Meters and Impedance Analyzers range from 1mHz to 3GHz devices to suit a wide range of applications in the testing of electronic components. The IM3533 series are Hioki’s first LCR meters to offer turn ration, mutual inductance and inductance difference measurements of transformers and coils, and are integrated with an internal DC bias to facilitate HDMI compliance testing.

KEY FEATURES

• ±0.05% accuracy with wide measurement range
(DC, 1mHz to 200kHz, 5mV to 5V, 10uA to 50mA)
• Non-stop testing over mixed measurement conditions such as C-D and ESR at 10 times the speed of previous models
• Built-in low impedance high precision mode effective for testing low inductance or the ESR of aluminum electrolysis capacitance
• Dedicated modes for measuring transformer winding ratio, mutual inductance and temperature compensated DCR
• Frequency sweep testing (IM3533-01 only)
• 2m/4m cable setting in addition to the standard 0m/1m
• Touch screen with intuitive operation

MODEL NO. (ORDER CODE)

IM3533
IM3533-01Advanced function model

Note: This product is not supplied with measurement probes or test fixtures. Please select and purchase the measurement probe or test fixture options appropriate for your application separately. All probes are constructed with a 1.5D-2V coaxial cable. For an RS-232C connection: A crossover cable for interconnection can be used. You can use the RS-232C CABLE 9637 without hardware flow control.

LAN
USB 2.0
CE
GPIB
RS232C
https://youtube.com/watch?v=WYD7DGCbmvI%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

LCR METER IM3533 MEASUREMENT METHODS: CONTINUOUS MEASUREMENT USING DIFFERENT SETS OF CONDITIONS

The Hioki IM3533 LCR Meter can perform continuous measurement using different capacitor measurement conditions, for example to perform C-D measurement at 120 Hz and ESR measurement at 100 kHz, without stopping to manually change settings.

https://youtube.com/watch?v=3giTLF-e_HA%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

LCR METER IM3533 MEASUREMENT METHODS: TURN RATIO MEASUREMENT

In transformer and winding measurement, the Hioki IM3533 LCR meter can measure primary-side and secondary-side L values and then calculate and display the turn ratio. This capability eliminates the need to manually calculate this value, as in the conventional approach.

https://youtube.com/watch?v=P0-iGnAW7FM%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

LCR METER IM3533 MEASUREMENT METHODS: MUTUAL INDUCTANCE MEASUREMENT

In transformer measurement, the Hioki IM3533 LCR Meter can perform L measurement with a same-phase series connection and again with a reverse-phase series connection, and then calculate and display the mutual inductance. This capability eliminates the need to manually calculate this value, making it more convenient and efficient than the conventional approach.

https://youtube.com/watch?v=RIHiYloj4hE%3Frel%3D0%26enablejsapi%3D1%26wmode%3Dopaque

LCR METER IM3533-01 MEASUREMENT METHODS: FREQUENCY SWEEP MEASUREMENT

The IM3533-01 can perform automatic measurement while sweeping through up to 801 frequencies from a user-specified frequency range or frequency list. Measurement results can be saved on a USB flash drive or computer via the instrument’s interface in order to create lists and graphs of frequency characteristics using the included LCR application software.

APPLICATION INFORMATION

Download the sample application and view the Communications Command User Manual.

Click here for the LCR application.

ACCURACY CALCULATION INFORMATION

Calculate accuracy by simply entering a numerical value.

Click here to calculate accuracy.

BASIC SPECIFICATIONS (ACCURACY GUARANTEED FOR 1 YEAR, POST-ADJUSTMENT ACCURACY GUARANTEED FOR 1 YEAR)

 IM3533IM3533-01
Measurement modesLCR, Transformer testing (N, M, ΔL), Continuous testing (LCR mode)LCR, Transformer testing (N, M, ΔL), Analyzer (sweep testing), Continuous Testing (LCR/Analyzer mode)
Measurement parametersZ, Y, θ, Rs (ESR), Rp, Rdc (DC resistance), X, G, B, Cs, Cp, Ls, Lp, D (tanδ), Q, N, M, ΔL, T
Measurement range100 mΩ to 100 MΩ, 10 ranges (All parameters defined in terms of Z.)
Displayable rangeZ, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp : ± (0.00000 [unit] to 9.99999G [unit]) Real value display for Z and Y only
θ: ± (0.000° to 180.000°), D: ± (0.00000 to 9.99999)
Q: ± (0.00 to 99999.9), Δ%: ± (0.0000% to 999.999%), T: -10.0°C to 99.9°C
Basic accuracyZ : ±0.05% rdg. θ: ±0.03°
Measurement frequency1 mHz to 200 kHz (5 digits setting resolution, minimum resolution 1 mHz)
Measurement signal level[Normal mode]
V mode, CV mode: 5 mV to 5 Vrms, 1 mVrms steps
CC mode: 10 μA to 50 mArms, 10 μArms steps
[Low impedance high accuracy mode]
V mode, CV mode: 5 mV to 2.5 Vrms, 1 mVrms steps
CC mode: 10 μA to 100 mArms, 10 μArms steps
Output impedanceNormal mode: 100 Ω, Low impedance high accuracy mode: 25 Ω
Display5.7-inch touch-screen color TFT, display can be set to ON/OFF
Measurement time2 ms (1 kHz, FAST, display OFF, representative value)
FunctionsDC bias measurement, DC resistance temperature compensation (converted reference temperature display), Comparator, BIN measurement (classify function), Panel loading/saving, Memory function
InterfacesEXT I/O (Handler), USB communication (high-speed), USB memory
Optional: Choose 1 from RS-232C, GP-IB, or LAN
Power supply100 to 240 V AC, 50/60 Hz, 50 VA max
Dimensions and mass330 mm (12.99 in) W × 119 mm (4.69 in) H × 168 mm (6.61 in) D, 3.1 kg (109.3 oz)
AccessoriesPower cord × 1, Instruction manual × 1, CD-R (Includes PC commands and sample software) × 1

CATALOGS

CATALOG: LCR METER SERIES 

DOWNLOAD

LCR METER IM3523, IM3533 

DOWNLOAD

SMD TEST FIXTURE IM9110 

DOWNLOAD

ELECTRONIC COMPONENT MEASUREMENT APPLICATIONS 

DOWNLOAD

DECLARATION OF CONFORMITY

LCR METER IM3533,IM3533-01 CHEMICAL IMPEDANCE ANALYZER IM3590 [121KB] 

DOWNLOAD

OPTIONS

PC communication

*The 9151-02 is only for the RM3545-01

LAN INTERFACE Z3002

GP-IB INTERFACE Z3000

GP-IB CONNECTOR CABLE 9151-02

2 m (6.56 ft) length

RS-232C INTERFACE Z3001

Probe and Test fixtures

4-TERMINAL PROBE 9500-10

Cable length 1 m (3.28 ft), DC to 200 kHz, impedance characteristics of 50 Ω, measurable conductor diameter: φ0.3 mm (0.01 in) to 2 mm (0.08 in)

SMD TEST FIXTURE IM9110

• Direct connection two-terminal measurement type • DC to 1 MHz • Measurable sample sizes: 008004 (EIA)

PINCHER PROBE L2001

Cable length 73 cm (2.40 ft), DC to 8 MHz, impedance characteristics of 50 Ω, 4-terminal pair configuration, tip electrode spacing: 0.3 (0.01 in) to 6 mm (0.24 in)

4-TERMINAL PROBE L2000

Cable length 1 m (3.28 ft), DC to 8 MHz, impedance characteristics of 50 Ω, 4-terminal pair configuration, measurable conductor diameter: ø0.3 (0.01 in) to 5 mm (0.20 in)

CONTACT TIPS IM9902

To replace the tip on the L2001, small size

CONTACT TIPS IM9901

To replace the tip on the L2001, regular size, bundled with the L2001

SMD TEST FIXTURE 9699

Direct connection type, For measuring SMDs with electrodes on the bottom; DC to 120 MHz, test sample dimensions: 1.0 mm (0.04 in) to 4.0 mm (0.16 in) wide, max. 1.5 mm (0.06 in) high

SMD TEST FIXTURE 9677

Direct connection type, For measuring SMDs with electrodes on the side; DC to 120 MHz, test sample dimensions: 3.5 mm ±0.5 mm (0.14 in ±0.02 in)

DC BIAS CURRENT UNIT 9269-10

Direct connection type, 40 Hz to 2 MHz, maximum applied current of DC 2 A

DC BIAS VOLTAGE UNIT 9268-10

Direct connection type, 40 Hz to 8 MHz, maximum applied voltage of DC ±40 V

SMD TEST FIXTURE 9263

Direct connection type, DC to 8 MHz, Test sample dimensions:1 mm (0.04 in) to 10 mm (0.39 in)

TEST FIXTURE 9262

Direct connection type, DC to 8 MHz, measurable conductor diameter: ø0.3 (0.01 in) to 2 mm (0.08 in)

TEST FIXTURE 9261-10

Cable length 1 m (3.28 ft), DC to 8 MHz, impedance characteristics of 50 Ω, 4-terminal pair configuration, measurable conductor diameter: ø0.3 (0.01 in) to 1.5 mm (0.06 in)

4-TERMINAL PROBE 9140-10

Cable length 1 m (3.28 ft), DC to 200 kHz, impedance characteristics of 50 Ω, 4-terminal pair configuration, measurable conductor diameter: ø0.3 (0.01 in) to 5 mm (0.20 in)

SMD TEST FIXTURE IM9100

• Direct connection type • SMDs with electrodes on the bottom • DC to 8MHz • Measurable sample sizes: 01005 to 0402 (EIA)

Temperature probe

SHEATH TYPE TEMPERATURE PROBE 9478

Pt100, Tip dia. φ2.3 mm (0.09 in), Cord length 1 m (3.28 ft), Waterproof construction